FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4233524
·
Received November 6, 2014
Report
- Report Number
- 2032227-2014-48792
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- February 19, 2014
- Report Date
- February 19, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PUMP PASSED DISPLACEMENT TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, EXCESSIVE NO DELIVERY TEST AND PRIME/A33 TEST. MOTOR PASSED MOTOR TEST. NO MOTOR ERROR ALARM. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW. UNIT RECEIVED CRACKED CASE DISPLAY WINDOW CORNER. UNIT RECEIVED WITH CRACKED BATTERY TUBE THREADS. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED A MOTOR ERROR ALARM. BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 712915 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LWWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |