FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4233524 · Received November 6, 2014

Report

Report Number
2032227-2014-48792
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
February 19, 2014
Report Date
February 19, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PUMP PASSED DISPLACEMENT TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, EXCESSIVE NO DELIVERY TEST AND PRIME/A33 TEST. MOTOR PASSED MOTOR TEST. NO MOTOR ERROR ALARM. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW. UNIT RECEIVED CRACKED CASE DISPLAY WINDOW CORNER. UNIT RECEIVED WITH CRACKED BATTERY TUBE THREADS. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED A MOTOR ERROR ALARM. BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
712915 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LWWL

Patients

Seq Age Sex Outcome Treatment
1