PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-48843
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 7, 2014
- Report Date
- October 8, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP ALARMED FOR A MOTOR ERROR DURING THE BASIC OCCLUSION TEST AND WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO FAULTY FORCE SENSOR RESISTOR. THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON THE DISPLAY WINDOW.
THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR MULTIPLE TIMES. NO FURTHER INFORMATION REPORTED.