FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4233472 · Received November 6, 2014

Report

Report Number
2032227-2014-48591
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
June 19, 2012
Report Date
October 13, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM NOTED. PUMP HAS INTERMITTENT BUTTON RESPONSE DUE TO CORRODED DAMAGE ON KEYPAD TRACES. PUMP RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED WINDOW AND CRACKED BATTERY TUBE THREADS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LED TO THE ALARM OR IF THE DEVICE WAS DROPPED. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 206 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
713144 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 46 YR