FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4233470
·
Received November 6, 2014
Report
- Report Number
- 2032227-2014-48590
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- May 15, 2011
- Report Date
- May 15, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
UNRESPONSIVE BUTTON DUE TO FLATTENED DOME SWITCH AT ACT BUTTON ON KEYPAD TRACE. ALSO, UNIT HAD MISSING END CAP STICKER.(B)(4)
Description of Event or Problem · 1
IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP KEYPAD WAS UNRESPONSIVE TO USER INPUT. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LED TO THE ALARM OR IF THE DEVICE WAS DROPPED. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 111 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.