PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-48578
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 6, 2014
- Report Date
- October 8, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACES. THE INSULIN PUMP HAD A CRACKED DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, SCRATCHED RESERVOIR TUBE WINDOW, CRACKED BATTERY TUBE THREADS AND A CRACKED BELT CLIP SLOT.
CUSTOMER REPORTED A BUTTON ERROR ON HER INSULIN PUMP. THE LAST FEW DAYS THE ESC BUTTON IS UNRESPONSIVE. CUSTOMER HAS TO PRESS THE BUTTON SEVERAL TIMES. THIS FIRST OCCURRED ON (B)(6) 2014. CUSTOMER'S BLOOD GLUCOSE IS 117 MG/DL. THERE WERE NO SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER WAS ADVISED THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 712772 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |