PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4233425 · Received November 6, 2014

Report

Report Number
2032227-2014-48517
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
October 8, 2014
Report Date
October 8, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACES. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, A BROKEN BELT CLIP SLOT, A CRACKED RESERVOIR TUBE LIP, A CRACKED RESERVOIR TUBE AND A MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 139MG/DL. TROUBLESHOOTING OCCURRED. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.