FENESTRATED BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-05624
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- June 5, 2014
- Report Date
- September 16, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND THE INSTRUMENT PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. THE REPORTED COMPLAINT OF A BROKEN WIRE DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT. RECURRENCE OF THE ALLEGED FAILURE MODE IS NOT EXPECTED TO BE LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH, SERIOUS INJURY OR SERIOUS DETERIORATION IN THE STATE OF HEALTH OF A PATIENT. HOWEVER, THE BROKEN PITCH CABLE FOUND DURING FAILURE ANALYSIS INVESTIGATIONS COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE MALFUNCTION WERE TO RECUR. INTUITIVE SURGICAL INC. (ISI) HAS CONDUCTED A DEVICE HISTORY RECORD (DHR) REVIEW FOR THIS DEVICE AND DID NOT FIND ANY NON-CONFORMANCES THAT WOULD AFFECT ANY MATERIAL OF THE FINAL PRODUCT AND/OR THE QUALITY OR PERFORMANCE OF THE INSTRUMENT.
IT WAS REPORTED DURING A DA VINCI SURGICAL PROCEDURE, THE WIRE BROKE ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 713908 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M10140206 321 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |