FDA Adverse Event Malfunction Summary report: N

COULTER LH 750 HEMATOLOGY ANALYZER

MDR report key: 4233349 · Received November 6, 2014

Report

Report Number
1061932-2014-02772
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
October 15, 2014
Report Date
October 15, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) DISCOVERED POOR CONNECTION AT THE LS PREAMP AND LS SENSOR. THE FSE REPLACED THE LS PREAMP AND LS SENSOR TO RESOLVE THE REPORTED ISSUE. THE FSE COMPLETED ALIGNMENT TO SPECIFICATION. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED LOW DIFF AND RETIC LATRON LIGHT SCATTER INVOLVING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE CUSTOMER STATED PATIENT RESULTS WERE NOT IMPACTED - NO ERRONEOUS RESULTS WERE GENERATED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
713386 COULTER LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1