FDA Adverse Event Malfunction Summary report: N

RUBY COIL

MDR report key: 4233316 · Received November 6, 2014

Report

Report Number
3005168196-2014-00798
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
May 28, 2014
Report Date
May 28, 2014
Manufacturer
PENUMBRA, INC.
Product Code
HCG
PMA / PMN Number
K120330
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION: THE PRODUCT WAS NOT RETURNED FOR EVALUATION. WITHOUT THE RETURN OF THE DEVICE, THE ROOT CAUSE OF THE PROBLEM CANNOT BE DETERMINED. THE MANUFACTURING RECORDS FOR THIS LOT WERE REVIEWED AND DID NOT REVEAL ANY OUTSTANDING DISCREPANCIES, DESIGN, OR QUALITY CONCERNS. THIS MDR IS ASSOCIATED WITH MDR 3005168196-2014-00796, 00797, AND 00799. DEVICE WAS DISPOSED OF BY HOSPITAL.

Description of Event or Problem · 1

THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING A PX SLIM DELIVERY MICROCATHETER AND RUBY COILS. DURING THE PROCEDURE, THE PHYSICIAN WAS UNABLE TO ADVANCE THE RUBY COIL INTO PX SLIM DELIVERY MICROCATHETER. THE PHYSICIAN REMOVED THE ENTIRE RUBY COIL FROM THE INTRODUCER AND RE-LOADED THE RUBY COIL INTO THE INTRODUCER; HOWEVER, THE RUBY COIL WOULD NOT ADVANCE SMOOTHLY THROUGH THE PX SLIM DELIVERY MICROCATHETER. SEVERAL ATTEMPTS WERE MADE TO FLUSH THE COIL INSIDE THE INTRODUCER SHEATH BEFORE TRYING TO PASS IT THOUGH PX SLIM DELIVERY MICROCATHETER; HOWEVER, THE RUBY COIL BUCKLED INSIDE THE ROTATING HEMOSTATIC VALVE (RHV). THE RUBY COIL WAS REMOVED AND ANOTHER RUBY COIL WAS USED. WHEN THE SECOND RUBY COIL WAS REMOVED ITS PACKAGE, A SIMILAR INCIDENT OCCURRED. THE RUBY COIL SEEMED WEDGED INSIDE THE INTRODUCER SHEATH AND UNABLE TO PASS THROUGH THE PX SLIM DELIVERY MICROCATHETER WITH HUGE RESISTANCE AND BUCKLING BACK INTO THE RHV. THE SECOND RUBY COIL WAS REMOVED AND ANOTHER RUBY COIL WAS USED. THE FIRST FEW CENTIMETERS OF THE THIRD RUBY COIL BEGAN TO PASS THROUGH PX SLIM DELIVERY MICROCATHETER; HOWEVER, A STICKY RESISTANCE FEELING OCCURRED. AS THE PHYSICIAN PULLED BACK ON THE RUBY COIL, IT UNINTENTIONALLY DETACHED INSIDE THE PX SLIM DELIVERY MICROCATHETER. THE RUBY COIL AND THE PX SLIM DELIVERY MICROCATHETER WERE REMOVED AND NEW ONES WERE USED. THERE WAS NO REPORT OF AN ADVERSE EFFECT ON THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
713312 RUBY COIL HCG, KRD HCG PENUMBRA, INC. F40043

Patients

Seq Age Sex Outcome Treatment
1