EMAX 2 PLUS MOTOR
Report
- Report Number
- 1045834-2014-14840
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 29, 2014
- Report Date
- October 29, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK080802
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE FAILED THERMISTOR ASSESSMENT AND THE ROTATIONS PER MINUTE (RPMS) WERE AT 49,000. IT WAS DETERMINED THAT THIS TYPE OF DAMAGE WAS INDICATIVE OF A DEVICE THAT HAD OVERHEATED, WHICH CAUSED THE THERMISTOR TO GO BAD AND NO LONGER OPERATE PROPERLY. IT WAS DETERMINED THAT THE LOW RPMS WERE DUE TO A THERMISTOR THAT WAS GOING OUT DUE TO OVERHEATING. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT THE MOTOR DEVICE WAS RUNNING HOT. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715552 | EMAX 2 PLUS MOTOR | MOTOR, DRILL, ELECTRIC - HANDPIECE | HBC | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |