PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4233112 · Received November 6, 2014

Report

Report Number
2032227-2014-48686
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
October 9, 2014
Report Date
October 9, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD A BLANK DISPLAY DUE TO A DISCONNECTED BATTERY TUBE. THE INSULIN PUMP HAD A DETACHED END CAP AND WAS MISSING THE MOTOR. THE INSULIN PUMP WAS ALSO RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, AND MINOR SCRATCHES ON THE DISPLAY WINDOW AND CASE.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.(B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP WAS DROPPED. THE BOTTOM END OF THE INSULIN PUMP BROKE OFF. THE PATIENT'S BLOOD GLUCOSE WAS 86 MG/DL. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714985 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 14 YR