530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-48640
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 9, 2014
- Report Date
- October 9, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHED LCD WINDOW.
IT WAS REPORTED THAT THE CUSTOMER HAD A MOTOR ERROR ALARM ON THE INSULIN PUMP. CUSTOMER HIT THE ESC BUTTON AND THE ALARM OCCURRED. CUSTOMER HAS ALSO GOTTEN A SENSOR ERROR. CUSTOMER USES THE SENSOR FEATURE ON THE PUMP. IT WAS EXPLAINED THAT A FALSE MOTOR ERROR MAY BE CAUSED BY VIEWING THE SENSOR GLUCOSE GRAPH WHILE THE PUMP IS DELIVERING A BOLUS. CUSTOMER STATED THAT SHE WAS ABLE TO COMPLETE THE REWIND SEQUENCE. CUSTOMER STATED THAT SHE WAS ABLE TO COMPLETE THE REWIND, BUT SHE IS STUCK IN SUSPEND. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROFESSIONAL'S INSTRUCTION. CUSTOMER WAS ADVISED TO RETURN THE PUMP WITH THE BATTERY AND ZERO OUT THE BASAL RATES. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715453 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-551NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |