DIMENSION VISTA CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2014-00568
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 13, 2014
- Report Date
- October 14, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJE
- PMA / PMN Number
- K051087
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER SITE. DURING EVALUATION OF THE INSTRUMENT, THE CSE OBSERVED CRUSTY RESIDUE ON THE ALIQUOT PROBE HOUSING. THE CSE REPLACED AND AUTO-ALIGNED THE ALIQUOT PROBE, SAMPLE PROBE 1, AND REAGENT PROBE 2 AND CLEANED DRAINS. THE CSE ALSO RAN A QUICK CHECK, WHICH PASSED, AND QUALITY CONTROLS, WHICH RESULTED WITHIN RANGE. THE CAUSE OF THE DISCORDANT RESULTS ON TWO PATIENT SAMPLES IS UNKNOWN. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.
DISCORDANT RESULTS WERE OBTAINED ON TWO PATIENT SAMPLES TESTED FOR MULTIPLE METHODS ON A DIMENSION VISTA 1500 INSTRUMENT. THE DISCORDANT RESULTS FOR THE FIRST PATIENT (B)(6) WERE REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED ON THE SAME INSTRUMENT AND THE REPEAT RESULTS WERE ISSUED TO THE PHYSICIAN(S). THE CUSTOMER STATED THAT THE SAMPLE WAS REPEATED ON AN ALTERNATE DIMENSION VISTA INSTRUMENT AND THE RESULTS MATCHED THE REPEAT RESULTS FROM THE ORIGINAL INSTRUMENT AND ALSO ALIGNED WITH RESULTS FROM TWO SAMPLES DRAWN FROM THE PATIENT LATER IN THE DAY. THE SECOND PATIENT SAMPLE (B)(6) WAS REPEATED ON THE SAME INSTRUMENT AND A DIMENSION EXL INSTRUMENT AND ALL REPEAT RESULTS MATCHED. IT IS UNKNOWN IF THE DISCORDANT OR REPEAT RESULTS WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715434 | DIMENSION VISTA CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC. | DIMENSION VISTA 1500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |