FDA Adverse Event Injury Summary report: N

530G INSULIN PUMP

MDR report key: 4233034 · Received November 6, 2014

Report

Report Number
2032227-2014-48657
Event Type
Injury
Date Received
November 6, 2014
Date of Event
October 9, 2014
Report Date
October 9, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ME
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD LOW BLOOD GLUCOSE LEVELS OF 42 MG/DL. THE CUSTOMER DECLINED TO TROUBLESHOOT THE LOW BLOOD GLUCOSE LEVELS. THE CUSTOMER REPORTED A METER BG NOW ALERT FROM THE INSULIN PUMP. THE CUSTOMER WAS ASSISTED IN CALIBRATING THE INSULIN PUMP. THE CUSTOMER HAD HIGH BLOOD GLUCOSE LEVELS OF 151 MG/DL AFTER TROUBLESHOOTING THE INSULIN PUMP. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715378 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-551NAS

Patients

Seq Age Sex Outcome Treatment
1 65 YR