FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4232998
·
Received November 6, 2014
Report
- Report Number
- 2032227-2014-48606
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- July 31, 2014
- Report Date
- July 31, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
UNIT RECEIVED WITH BLANK DISPLAY DUE TO MOISTURE DAMAGE AT ELECTRONIC ASSEMBLY. UNIT RECEIVED MOISTURE DAMAGED AT MOTOR ASSEMBLY. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW. UNIT RECEIVED WITH CRACKED BATTERY TUBE THREADS. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE INSULIN PUMP HAD VISIBLE MOISTURE UNDER THE DISPLAY SCREEN AFTER BEING EXPOSED TO STEAM FROM THE SHOWER. BLOOD GLUCOSE LEVEL WAS 252 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715283 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR |