FDA Adverse Event
Malfunction
Summary report: N
V200 VENTILATOR
MDR report key: 4232973
·
Received November 6, 2014
Report
- Report Number
- 2031642-2014-01373
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Report Date
- October 16, 2014
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- CBK
- PMA / PMN Number
- K102054
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE INTERNATIONAL CUSTOMER REPORTED THE VENTILATOR COULD NOT BOOT UP. THE CUSTOMER REPORTED THE DEVICE WAS NOT IN USE THEREFORE THERE WAS NO PATIENT INVOLVEMENT. THE MANUFACTURERS SERVICE PROVIDER CONFIRMED THE REPORTED PROBLEM. THE SERVICE PROVIDER REPLACED THE POWER SUPPLY TO ADDRESS THE REPORTED PROBLEM. POWER FAILURE DUE TO LOSS OF AC OR BATTERY POWER MAY RESULT IN SHUTDOWN OF DEVICE DURING NORMAL VENTILATION OPERATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715739 | V200 VENTILATOR | VENTILATOR, CONTINUOUS | CBK | RESPIRONICS CALIFORNIA, INC. | V200 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |