75MM SELECTABLE NEW TLC
Report
- Report Number
- 3005075853-2014-07694
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- June 17, 2014
- Report Date
- June 23, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K092577
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE NTLC75 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AS THE FIRING KNOB WAS DETACHED AND MISSING AND THE SLIP BLOCK ASSEMBLY WAS NOTED TO BE DAMAGED. THE DEVICE WAS RECEIVED WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RETURNED FULLY FIRED. DUE TO THE CONDITION OF THE DEVICE, NO FUNCTIONAL TEST COULD BE PERFORMED. THE DAMAGE TO THE FIRING KNOB AND SLIP BLOCK ASSEMBLY IS CONSISTENT WITH HIGH (OUTSIDE INDICATED USE) STAPLE FORMING FORCES; HOWEVER THERE IS INSUFFICIENT EVIDENCE TO DETERMINE THE CAUSE OF THE HIGHER LOADS. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A CONVENTIONAL OPEN BARIATRIC PROCEDURE, DURING GASTRIC STAPLING, THE THIRD BLUE RELOAD, THERE WAS LOCKING THAT FORCED THE CURSOR; IT'S BREAKDOWN OCCURRED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 714619 | 75MM SELECTABLE NEW TLC | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | K4DR1R |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | BRKBANTLC KIT |