FDA Adverse Event Malfunction Summary report: N

75MM SELECTABLE NEW TLC

MDR report key: 4232953 · Received November 6, 2014

Report

Report Number
3005075853-2014-07694
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
June 17, 2014
Report Date
June 23, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K092577
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE NTLC75 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AS THE FIRING KNOB WAS DETACHED AND MISSING AND THE SLIP BLOCK ASSEMBLY WAS NOTED TO BE DAMAGED. THE DEVICE WAS RECEIVED WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RETURNED FULLY FIRED. DUE TO THE CONDITION OF THE DEVICE, NO FUNCTIONAL TEST COULD BE PERFORMED. THE DAMAGE TO THE FIRING KNOB AND SLIP BLOCK ASSEMBLY IS CONSISTENT WITH HIGH (OUTSIDE INDICATED USE) STAPLE FORMING FORCES; HOWEVER THERE IS INSUFFICIENT EVIDENCE TO DETERMINE THE CAUSE OF THE HIGHER LOADS. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CONVENTIONAL OPEN BARIATRIC PROCEDURE, DURING GASTRIC STAPLING, THE THIRD BLUE RELOAD, THERE WAS LOCKING THAT FORCED THE CURSOR; IT'S BREAKDOWN OCCURRED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714619 75MM SELECTABLE NEW TLC STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA K4DR1R

Patients

Seq Age Sex Outcome Treatment
1 BRKBANTLC KIT