FDA Adverse Event Malfunction Summary report: N

V200 VENTILATOR

MDR report key: 4232949 · Received November 6, 2014

Report

Report Number
2031642-2014-01375
Event Type
Malfunction
Date Received
November 6, 2014
Report Date
October 16, 2014
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K102054
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE INTERNATIONAL CUSTOMER REPORTED THE VENTILATOR COULD NOT BOOT UP. THE CUSTOMER REPORTED THE DEVICE WAS NOT IN USE THEREFORE THERE WAS NO PATIENT INVOLVEMENT. THE MANUFACTURERS SERVICE PROVIDER CONFIRMED THE REPORTED PROBLEM. THE SERVICE PROVIDER REPLACED THE POWER SUPPLY TO ADDRESS THE REPORTED PROBLEM. POWER FAILURE DUE TO LOSS OF AC OR BATTERY POWER MAY RESULT IN SHUTDOWN OF DEVICE DURING NORMAL VENTILATION OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715706 V200 VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V200 NA

Patients

Seq Age Sex Outcome Treatment
1