FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 4232761 · Received September 12, 2014

Report

Report Number
1824206-2014-02245
Event Type
Malfunction
Date Received
September 12, 2014
Date of Event
August 15, 2014
Report Date
August 15, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

HILL-ROM TECHNICAL SUPPORT FOUND THE SCALE BOARD ASSEMBLY BAD. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE SCALE BOARD ASSEMBLY TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED EXIT ALARM IS INOPERABLE. THE BED WAS LOCATED AT THE ACCOUNT IN 4 WEST. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
564846 VERSACARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment
1