FDA Adverse Event
Malfunction
Summary report: N
CAREASSIST BED
MDR report key: 4232622
·
Received September 11, 2014
Report
- Report Number
- 3006697241-2014-00662
- Event Type
- Malfunction
- Date Received
- September 11, 2014
- Date of Event
- August 14, 2014
- Report Date
- August 14, 2014
- Manufacturer
- HILLROM DE MEXICO S DE RL DE CV
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE COMMUNICATION CABLE HAD A BROKEN PIN. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012-2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE COMMUNICATION CABLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED IS OT COMMUNICATING WITH THE NURSE CALL STATION. THE BED WAS LOCATED AT THE ACCOUNT IN THE BAILEY CENTER RM. 114. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #181192(B)(4)
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 563107 | CAREASSIST BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILLROM DE MEXICO S DE RL DE CV | 1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |