FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 4232622 · Received September 11, 2014

Report

Report Number
3006697241-2014-00662
Event Type
Malfunction
Date Received
September 11, 2014
Date of Event
August 14, 2014
Report Date
August 14, 2014
Manufacturer
HILLROM DE MEXICO S DE RL DE CV
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE COMMUNICATION CABLE HAD A BROKEN PIN. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012-2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE COMMUNICATION CABLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED IS OT COMMUNICATING WITH THE NURSE CALL STATION. THE BED WAS LOCATED AT THE ACCOUNT IN THE BAILEY CENTER RM. 114. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #181192(B)(4)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
563107 CAREASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILLROM DE MEXICO S DE RL DE CV 1170

Patients

Seq Age Sex Outcome Treatment
1