FDA Adverse Event
Malfunction
Summary report: N
SAM JUNCTIONAL TOURNIQUET (SJT)
MDR report key: 4232501
·
Received October 24, 2014
Report
- Report Number
- 3023316-2014-00003
- Event Type
- Malfunction
- Date Received
- October 24, 2014
- Date of Event
- September 1, 2014
- Report Date
- September 23, 2014
- Manufacturer
- SAM MEDICAL PRODUCTS
- Product Code
- DXC
- PMA / PMN Number
- K131561
- Removal / Correction Number
- 3023316-09/25/2014-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE FAILURE WAS CONCLUDED TO BE DUE TO A MANUFACTURING FLAW IN THE BROKEN SNAP HOOK.
Description of Event or Problem · 1
THE PLASTIC HOOK (SNAP HOOK) FROM THE AUXILIARY STRAP BROKE WHILE THE SJT WAS BEING APPLIED, AS PART OF A SIMULATED TRAINING EXERCISE, TO A VOLUNTEER. IT WAS NOT USED IN A CLINICAL SETTING OR APPLIED TO A PATIENT, SO THERE WAS NO POTENTIAL FOR HARM OR COMPROMISE TO TREATMENT IN THIS CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 677969 | SAM JUNCTIONAL TOURNIQUET (SJT) | JUNCTIONAL TOURNIQUET | DXC | SAM MEDICAL PRODUCTS | 102-0175-00 | 03/31/2014 |