FDA Adverse Event Injury Summary report: N

TELIGEN

MDR report key: 4232415 · Received November 6, 2014

Report

Report Number
2124215-2014-19596
Event Type
Injury
Date Received
November 6, 2014
Date of Event
September 18, 2014
Report Date
September 27, 2016
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Removal / Correction Number
Z-0088-2015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION WAS RECEIVED INDICATING A REPLACEMENT PROCEDURE WAS PLANNED WITH ANOTHER MANUFACTURER. RECORDS INDICATE THIS DEVICE REMAINS IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

(B)(4). RECORDS INDICATE THIS DEVICE REMAINS IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

THIS DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, OR IF ANALYSIS IS COMPLETED, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EMITTED BEEPING TONES. INTERROGATION REVEALED THE DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. A DEVICE REPLACEMENT WAS SCHEDULED WITH ANOTHER MANUFACTURER. NO ADVERSE PATIENT EFFECTS WERE REPORTED. BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO A POSSIBLE LOW VOLTAGE CAPACITOR ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED INDICATING THIS DEVICE WAS EXPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715882 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWP GUIDANT CRM CLONMEL IRELAND E110

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization| L| R 4136| E110