COGNIS
Report
- Report Number
- 2124215-2014-19521
- Event Type
- Injury
- Date Received
- November 6, 2014
- Date of Event
- August 19, 2014
- Report Date
- August 27, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Removal / Correction Number
- Z-0025-2014
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. REVIEW OF THE DEVICE MEMORY INDICATED THAT A LOW VOLTAGE ALERT, CODE 1003, WAS RECORDED. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH DEVICE PERFORMANCE WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE'S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE'S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE'S BATTERY FASTER THAN NORMAL. BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EMITTED BEEPING TONES AND WAS FOUND TO BE AT ELECTIVE REPLACEMENT INDICATOR (ERI). THERE WAS A CONCERN THE DEVICE WAS DEPLETING PREMATURELY MORE QUICKLY THAN EXPECTED. DISK ANALYSIS SHOWED THE BATTERY MAY BE DEPLETING MORE QUICKLY THAN EXPECTED AND DEVICE REPLACEMENT WAS RECOMMENDED. AN INVASIVE PROCEDURE WAS PERFORMED. THE DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715251 | COGNIS | IMPLANTABLE CHF PULSE GENERATOR | LWS | GUIDANT CRM CLONMEL IRELAND | N119 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Hospitalization| L| R | 0158| 4592| N119| MISMATCH| 4543| 1861| 4244 |