FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4232280 · Received November 6, 2014

Report

Report Number
2531779-2014-31751
Event Type
Malfunction
Date Received
November 6, 2014
Report Date
October 21, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/21/2014 WITH THE FOLLOWING FINDINGS: DURING A VISUAL INSPECTION OF THE PUMP, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED. DURING TESTING, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DIM AND DISCOLORED. THE OK KEYPAD BUTTON WAS INTERMITTENTLY RESPONSIVE TO USER PRESSES; ALL OTHER KEYPAD BUTTONS FUNCTIONED PROPERLY. THE KEYPAD COVER WAS REMOVED, AND CONTAMINATION WAS FOUND UNDER ALL BUTTON CONTACTS. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A CRACKED BATTERY COMPARTMENT, A DIM AND DISCOLORED DISPLAY, A KEYPAD BUTTON RESPONSE ISSUE, AND CONTAMINATION UNDER KEYPAD BUTTON CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/21/2014.