PROXIMATE ILS CIRCULAR STAPLER
Report
- Report Number
- 3005075853-2014-07668
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 17, 2014
- Report Date
- October 20, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K983536
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE CDH33A DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT BEFORE FIRING THE DEVICE THE ORANGE INDICATOR SHOULD BE FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE. IN ADDITION IT SHOULD BE NOTED THAT IF THE FIRING SEQUENCE IS NOT COMPLETE (THE FIRING HANDLE REACHES ITS STOPPING POINT, AND THE FIRING TRIGGER IS PARALLEL TO THE INSTRUMENT HANDLE) STAPLES COULD BE PARTIALLY DEPLOYED WITHOUT CUTTING THE WASHER. PLEASE REFERENCE THE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WAS MET PRIOR TO THE RELEASE OF THE DEVICE.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC ASSISTED SIGMOIDECTOMY/RIGHT HEMICOLECTOMY PROCEDURE, THE DEVICE STAPLED, BUT FAILED TO CUT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 713008 | PROXIMATE ILS CIRCULAR STAPLER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | L4E666 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |