FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4232022
·
Received November 6, 2014
Report
- Report Number
- 2032227-2014-48459
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 8, 2014
- Report Date
- October 8, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. THE INSULIN PUMP HAS MINOR SCRATCHED DISPLAY WINDOW AND CRACKED CASE AT DISPLAY WINDOW CORNERS.
Description of Event or Problem · 1
IT WAS REPORTED TAT THE CUSTOMER'S INSULIN PUMP WILL NOT POWER ON FOLLOWING A BATTERY CHANGE. HER BLOOD GLUCOSE WAS 121MG/DL AT THE TIME OF REPORTING. THE INSULIN PUMP IS BEING RETURNED. NO ADDITIONAL INFORMATION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 714694 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |