FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4232015 · Received November 6, 2014

Report

Report Number
2032227-2014-48451
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
July 23, 2014
Report Date
October 9, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FINDINGS: NO BUTTON ERROR ALARM NOTED. ACT BUTTON DID NOT RESPOND DUE TO CORRODED KEYPAD TRACE. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED DISPLAY WINDOW, CRACKED DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD A BUTTON ERROR THAT COULD NO BE CLEARED. HER BLOOD GLUCOSE WAS 57MG/DL AT THE TIME OF REPORTING. SHE ALSO STATED THAT THE BOTTOM OF THE SCREEN APPEARED TO HAVE A SLIGHT CRACK. THE INSULIN PUMP WAS RETURNED. NO ADDITIONAL INFORMATION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715770 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 54 YR