FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4231994 · Received November 6, 2014

Report

Report Number
2032227-2014-48450
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
May 3, 2013
Report Date
May 3, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM WAS NOTED. THE ESCAPE BUTTON DID NOT RESPOND DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP HAD A BUTTON ERROR THAT COULD NOT BE CLEARED. HIS OR HER BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF REPORTING. THE INSULIN PUMP IS BEING RETURNED. NO ADDITIONAL INFORMATION WAS REPORTED.