FDA Adverse Event Malfunction Summary report: N

XMAX MOTOR

MDR report key: 4231915 · Received November 6, 2014

Report

Report Number
1045834-2014-14764
Event Type
Malfunction
Date Received
November 6, 2014
Report Date
October 21, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
ERL
PMA / PMN Number
PK965080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE LOCKING MECHANISM WAS DAMAGED AND THE DEVICE COULD NOT SECURE A CUTTER BECAUSE IT WAS POWERBROKEN. IT WAS DETERMINED THAT THIS TYPE OF DAMAGE WITH PROLONGED USE WOULD MOST LIKELY CAUSE HEAT. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO FAILURE TO FOLLOW THE DIRECTIONS FOR USE AND RUNNING THE DEVICE IN THE SAFE OR LOAD POSITION. THIS WAS FURTHER DEFINED AS MISUSE, ABUSE AND/OR USER ERROR. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(6).AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT THE MOTOR DEVICE HAD ¿HEAT DURING OPERATION¿. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. SEVERAL ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715467 XMAX MOTOR DRILL, SURGICAL, ENT (PNEUMATIC) - HANDPIECE ERL DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1