RADICAL-7 HANDHELD
Report
- Report Number
- 2031172-2014-00332
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 19, 2014
- Report Date
- September 19, 2014
- Manufacturer
- MASIMO CORPORATION
- Product Code
- DQA
- PMA / PMN Number
- K110028
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE RETURNED DEVICE WAS EVALUATED. DURING EVALUATION THE DEVICE PASSED ALL VISUAL AND FUNCTIONAL TESTING. THE UNIT WAS FOUND TO VISUALLY AND AUDIBLY ALARM DURING ALARM CONDITIONS. TOUCHSCREEN WAS WORKING PROPERLY. THE UNIT WAS DETERMINED TO BE FUNCTIONING AS DESIGNED. A SERVICE HISTORY RECORD REVIEW REVEALS THAT THIS UNIT WAS IN THE FIELD FOR OVER NINE (9) MONTHS WITH NO PREVIOUS REPORTED ISSUES PRIOR TO THIS REPORTED EVENT.
MULTIPLE ATTEMPTS FOR PRODUCT RETURN AND REQUESTS FOR ADDITIONAL INFO WERE MADE. THE UNIT HAS NOT BEEN RETURNED TO MASIMO TO ALLOW AN ANALYSIS TO BE PERFORMED. IF NEW INFO IS OBTAINED OR THE DEVICE IS RETURNED, A FOLLOW UP REPORT WILL BE SUBMITTED.
THE CUSTOMER REPORTED THAT THE RADICAL-7 GOES INTO THE SPO2 SETTINGS SCREEN IMMEDIATELY AFTER ATTEMPTING TO GO TO THE HOME SCREEN. THE UNIT ENGAGES IN MONITORING. NO PT INCIDENT WAS REPORTED. ADDITIONAL INFO WAS RECEIVED ON 10/13/2014. AFTER POWERING UP, THE UNIT WOULD GO TO THE HOME SCREEN AND, AFTER A FEW SECONDS, IT WOULD CHANGE TO THE MAIN MENU SCREEN WITHOUT PHYSICAL CONTACT TO THE TOUCH SCREEN. AFTER USING THE HOME BUTTON, THE DEVICE WOULD GO TO THE HOME SCREEN AND, A FEW SECONDS LATER, WOULD RETURN TO THE MAIN MENU SCREEN WITHOUT ANY PHYSICAL CONTACT WITH THE TOUCH SCREEN. THIS CYCLE OCCURRED REPEATEDLY. THE CUSTOMER DIDN OT KNOW IF THE SETTINGS REMAINED UNCHANGED BECAUSE SHE DID NOT CHECK FOR THIS WHEN SHE EXPERIENCED THE REPORTED COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 662128 | RADICAL-7 HANDHELD | OXIMETER | DQA | MASIMO CORPORATION | 23785 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |