FDA Adverse Event Injury Summary report: N

840 VENTILATOR

MDR report key: 4231435 · Received November 6, 2014

Report

Report Number
8020893-2014-02448
Event Type
Injury
Date Received
November 6, 2014
Date of Event
October 22, 2014
Report Date
October 23, 2014
Manufacturer
COVIDIEN
Product Code
CBK
PMA / PMN Number
K970460
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE DEVICE, AND VERIFIED THE REPORTED MALFUNCTION. THE CSE REPLACED THE REPLACED THE ANALOG INTERFACE (AI) PRINTED CIRCUIT BOARD (PCB), AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MANUFACTURING SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY A CUSTOMER IN THE (B)(6) THAT, DURING PATIENT USE, THE 840 VENTILATOR SHUT DOWN. THE PATIENT WAS TRANSFERRED TO ANOTHER VENTILATOR. THERE WAS NO REPORTED PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714012 840 VENTILATOR CONTINUOUS VENTILATOR CBK COVIDIEN 840

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention