PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-48480
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 8, 2014
- Report Date
- October 8, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE DOWN ARROW BUTTON DUE TO CORRODED KEYPAD TRACE. THE INSULIN PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.
THE CUSTOMER REPORTED THAT THEIR KEYPAD WAS UNRESPONSIVE. THERE WAS NO SIGNIFICANT EVENTS LEADING TO THE KEYPAD ISSUES. THE CUSTOMER'S BLOOD GLUCOSE WAS 65 MG/DL. ADVISED CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO THEIR BACK-UP PLAN. NO FURTHER INFORMATION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715967 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |