FDA Adverse Event Malfunction Summary report: N

V60 VENTILATOR

MDR report key: 4231276 · Received November 6, 2014

Report

Report Number
2031642-2014-01368
Event Type
Malfunction
Date Received
November 6, 2014
Report Date
October 14, 2014
Manufacturer
REPIRONICS CALIFORNIA, INC
Product Code
MNT
PMA / PMN Number
K082660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

DURING DEVICE SERVICE, THE MANUFACTURER'S FIELD SERVICE ENGINEER (FSE) REPORTED REVIEW OF THE DEVICE DIAGNOSTIC HISTORY NOTED CODES INDICATING A PRESSURE SENSORS/SUPPLY FAILURE. THE FAILURES AS NOTED MAY RESULT IN VENT INOP/SHUTDOWN OF THE DEVICE DURING NORMAL VENTILATION OPERATION. THE FSE REPLACED THE CPU AND MOTOR CONTROLLER PCB BOARDS TO ADDRESS THE FINDING. FACTORY ANALYSIS OF THE MOTOR CONTROLLER PCB BOARD REVEALED A CAPACITOR FAILURE THAT RESULTED IN THE NOTED POWER ISSUE. TESTING OF THE CPU PCB BOARD IDENTIFIED NO FAILURES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714339 V60 VENTILATOR VENTILATOR, CONTINUOUS MNT REPIRONICS CALIFORNIA, INC V60 NA