FDA Adverse Event Malfunction Summary report: N

TENDRIL ST

MDR report key: 4231262 · Received November 6, 2014

Report

Report Number
2017865-2014-18082
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
August 25, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4).EVALUATION DESCRIPTION INCLUDED. NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND INSULATION DEGRADATION AT 4.5CM FROM THE CONNECTOR PIN. (B)(4).

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715362 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC., CRMD 1888TC/46 NA