PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-48320
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 7, 2014
- Report Date
- October 7, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
THE INSULIN PUMP ALARMED A33 DURING PRIME/A33 TEST AT 4 POUNDS DUE TO FAULTY FORCE SENSOR RESISTOR GOLD. THE INSULIN PUMP HAS MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS.
IT WAS REPORTED THAT THE CUSTOMER HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM ON THE INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 117 MG/DL. CUSTOMER STATED THE DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER DID NOT PRESS ON CAP. IT WAS EXPLAINED THAT DURING SEATING, THE FORCE SENSOR DID NOT DETECT THE RESERVOIR. INSULIN PUMP WILL NEED TO BE REPLACED. CUSTOMER WAS ADVISED TO REVERT TO A BACK-UP PLAN PER HEALTH CARE PROFESSIONAL'S INSTRUCTIONS. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 713733 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |