ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-31717
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Report Date
- October 15, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/15/2014 WITH THE FOLLOWING FINDINGS: THE DISPLAY WAS OBSERVED TO BE DIM, FADING, AND REDDISH IN COLOR. THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED BELOW THE BUMPER PAD. THE BLACK BOX SHOWED A TIME AND DATE RESET TO THE DEFAULT SETTING AFTER A BATTERY CHANGE ON (B)(6) 2014. THE PUMP WAS EXERCISE FOR 24 HOURS. THE BATTERY WAS THEN REMOVED FOR 6 HOURS AND A TIME AND DATE RESET OCCURRED WHEN THE BATTERY WAS REINSERTED. THE PUMP CASE WAS REMOVED FOR FURTHER INVESTIGATION AND CORROSION WAS FOUND UNDERNEATH THE INTERNAL CLOCK BATTERY COMPONENT. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DIM, FADING, REDDISH COLORED DISPLAY, A CRACKED BATTERY COMPARTMENT AND CORROSION UNDER THE INTERNAL CLOCK BATTERY COMPONENT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/15/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715525 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |