FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4230721 · Received November 6, 2014

Report

Report Number
2531779-2014-31717
Event Type
Malfunction
Date Received
November 6, 2014
Report Date
October 15, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/15/2014 WITH THE FOLLOWING FINDINGS: THE DISPLAY WAS OBSERVED TO BE DIM, FADING, AND REDDISH IN COLOR. THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED BELOW THE BUMPER PAD. THE BLACK BOX SHOWED A TIME AND DATE RESET TO THE DEFAULT SETTING AFTER A BATTERY CHANGE ON (B)(6) 2014. THE PUMP WAS EXERCISE FOR 24 HOURS. THE BATTERY WAS THEN REMOVED FOR 6 HOURS AND A TIME AND DATE RESET OCCURRED WHEN THE BATTERY WAS REINSERTED. THE PUMP CASE WAS REMOVED FOR FURTHER INVESTIGATION AND CORROSION WAS FOUND UNDERNEATH THE INTERNAL CLOCK BATTERY COMPONENT. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DIM, FADING, REDDISH COLORED DISPLAY, A CRACKED BATTERY COMPARTMENT AND CORROSION UNDER THE INTERNAL CLOCK BATTERY COMPONENT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/15/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715525 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR