FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4230716 · Received November 6, 2014

Report

Report Number
1416980-2014-39086
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
October 13, 2014
Report Date
October 13, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER OF THE DEVICE WAS UNKNOWN; THEREFORE, A SAMPLE ANALYSIS AND/OR A BATCH REVIEW COULD NOT BE COMPLETED. AFTER REVIEW, THE CAUSE OF THE REPORTED EVENT WAS UNABLE TO BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN SET/LINE) ALARM OCCURRED ON THE HOMECHOICE DEVICE DURING DWELL THREE OF FIVE. THE HOME PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. DURING TROUBLESHOOTING, THERE WAS NOTHING UNUSUAL NOTED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE REVIEWED PROPER PROCEDURES WITH THE PATIENT, EXPLAINED THE ALARM TO THE PATIENT, ASSISTED THE PATIENT TO CLEAR THE ALARM, AND ADVISED THE PATIENT TO BEGIN THERAPY AGAIN WITH ALL NEW SUPPLIES OR RESUME THERAPY WITH MANUAL BAGS. THE HOME PATIENT WOULD COMPLETE THERAPY USING MANUAL EXCHANGE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
716000 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 87 YR HOMECHOICE