FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4230700 · Received November 6, 2014

Report

Report Number
2032227-2014-26394
Event Type
Injury
Date Received
November 6, 2014
Date of Event
October 8, 2014
Report Date
October 8, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE CUSTOMER WAS ATTEMPTING A 5.9 UNIT CORRECTION WHEN HE RECEIVED A FINISHED LOADING ALARM. CUSTOMER'S BLOOD GLUCOSE WAS 500 MG/DL. HE STATED HE HAD TREATED HIS BLOOD GLUCOSE WITH A SYRINGE. WHILE ASSISTING THE CUSTOMER WITH CLEARING THE ALARM, IT WAS FOUND THE CUSTOMER HAD A PRIME/REWIND ANOMALY. CUSTOMER STATED HE WAS STUCK IN THE REWIND COMPLETE/BOLUS DELIVERY LOOP. CUSTOMER WAS ADVISED TO REMOVE HIS BATTERY. TROUBLESHOOTING WAS ABLE TO HELP THE CUSTOMER EXIT THE REWIND COMPLETE/BOLUS DELIVERY LOOP. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715409 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 62 YR