FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4230589 · Received November 6, 2014

Report

Report Number
1416980-2014-39074
Event Type
Malfunction
Date Received
November 6, 2014
Date of Event
October 13, 2014
Report Date
October 13, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE SAMPLE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, CONNECTING TO THE PATIENT LINE BEFORE IT IS PROPERLY PRIMED IS A KNOWN ERROR. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. THE GUIDE WARNS THE USER NOT TO CONNECT TO YOUR PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. THE GUIDE INSTRUCTS THE USER TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED. THE GUIDE ALSO PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY REPRIMING THE PATIENT LINE. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING TROUBLESHOOTING FOR AN UNRELATED ALARM ON THE HOMECHOICE MACHINE, IT WAS REPORTED THE HOME PATIENT HAD CONNECTED TO THE PATIENT LINE PRIOR TO COMPLETING THE PRIMING PHASE OF PERITONEAL DIALYSIS THERAPY. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED THE HOME PATIENT IN ENDING THERAPY AND REVIEWED PROPER PROCEDURES WITH THE HOME PATIENT. THE HOME PATIENT WOULD COMPLETE THERAPY USING ALL NEW SUPPLIES. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715087 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 65 YR HOMECHOICE