SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2014-39074
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 13, 2014
- Report Date
- October 13, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AS THE SAMPLE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. HOWEVER, CONNECTING TO THE PATIENT LINE BEFORE IT IS PROPERLY PRIMED IS A KNOWN ERROR. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. THE GUIDE WARNS THE USER NOT TO CONNECT TO YOUR PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. THE GUIDE INSTRUCTS THE USER TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED. THE GUIDE ALSO PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY REPRIMING THE PATIENT LINE. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING TROUBLESHOOTING FOR AN UNRELATED ALARM ON THE HOMECHOICE MACHINE, IT WAS REPORTED THE HOME PATIENT HAD CONNECTED TO THE PATIENT LINE PRIOR TO COMPLETING THE PRIMING PHASE OF PERITONEAL DIALYSIS THERAPY. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED THE HOME PATIENT IN ENDING THERAPY AND REVIEWED PROPER PROCEDURES WITH THE HOME PATIENT. THE HOME PATIENT WOULD COMPLETE THERAPY USING ALL NEW SUPPLIES. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715087 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | HOMECHOICE |