FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4230588 · Received November 6, 2014

Report

Report Number
2531779-2014-31709
Event Type
Malfunction
Date Received
November 6, 2014
Report Date
November 3, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/03/2014 WITH THE FOLLOWING FINDINGS: INVESTIGATORS PERFORMED REWIND, LOAD AND PRIME STEPS WITH NO ALARMS. THE PUMP WAS OPENED AND INSPECTED THE LEAD SCREW AND MOTOR DRIVE ASSEMBLY AND THERE WAS NO DEFECT FOUND. THE BATTERY COMPARTMENT WAS CRACKED FROM THE CASE SEAL TO THE GRIP. THERE WAS NO MOISTURE IN THE PUMP. THE DISPLAY WAS FADED AND PINK. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED THE DISPLAY WAS DIM AND PINK AND THE BATTERY COMPARTMENT WAS CRACKED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 11/03/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714587 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 35 YR