PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-37093
- Event Type
- Malfunction
- Date Received
- October 11, 2014
- Date of Event
- September 12, 2014
- Report Date
- September 12, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP HAD A BLANK DISPLAY DUE TO A BROKEN SOLDER JOINT ON THE INTERFACE BOARD. THE INSULIN PUMP WAS ALSO RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED RESERVOIR TUBE LIP, AND CRACKED BATTERY TUBE THREADS. NO FAILED BATTERY TEST ALARM COULD BE CONFIRMED DUE TO THE BLANK DISPLAY.
CUSTOMER REPORTS TO HAVE A BLANK DISPLAY SCREEN. CUSTOMER'S BLOOD GLUCOSE WAS 129 MG/DL. AFTER TROUBLESHOOTING, DISPLAY SCREEN DID NOT RETURN. CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. NO FURTHER INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 645682 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | MEDTRONIC MINIMED | MMT-723RNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |