FDA Adverse Event Malfunction Summary report: N

AED PLUS

MDR report key: 4230339 · Received October 21, 2014

Report

Report Number
1220908-2014-02661
Event Type
Malfunction
Date Received
October 21, 2014
Report Date
October 2, 2014
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K011541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY TRANSISTOR ON THE MAIN BOARD. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A BIOMED TESTING THE DEVICE INAPPROPRIATELY PROMPTED "UNIT FAILED" AND "CHANGE BATTERIES" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
668026 AED PLUS DEFIBRILLATOR MKJ ZOLL MEDICAL CORPORATION AED PLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA