SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2014-39073
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Date of Event
- October 13, 2014
- Report Date
- October 13, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS IS A REPORT OF A PATIENT WHO EXPERIENCED A AIR BUBBLES IN THE PATIENT LINE CAUSED BY THE PATIENT CONNECTING THEMSELVES AND INITIATING THERAPY PRIOR TO PROPERLY PRIMING THE PATIENT LINE. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. THE GUIDE WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THERE WERE AIR BUBBLES IN THE PATIENT LINE DURING INITIAL DRAIN OF PERITONEAL DIALYSIS THERAPY ON THE HOMECHOICE DEVICE. THE PATIENT WAS CONNECTED TO THE DEVICE AT THE TIME OF THE EVENT. DURING TROUBLESHOOTING, IT WAS DETERMINED THAT THE PATIENT HAD NOT COMPLETELY PRIMED THE PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED IN ENDING THE THERAPY. THE PATIENT PLANNED TO RESTART THERAPY USING NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 712732 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | HOMECHOICE |