FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4230314
·
Received November 6, 2014
Report
- Report Number
- 1031452-2014-17723
- Event Type
- Malfunction
- Date Received
- November 6, 2014
- Report Date
- October 16, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER INDEPENDENT REPAIR CENTER, THE UNIT HAD LOW O2 OR YELLOW LIGHT. THE KEY FAILURE WAS THE SIEVE BEDS BEING BROKEN. ADDITIONAL MALFUNCTION FOR THIS DEVICE WAS THE PC BOARD HAVING A SHORT CIRCUIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 713705 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |