FDA Adverse Event
Malfunction
Summary report: N
MEDFUSTION 3500 SYRINGE INFUSION PUMP
MDR report key: 4230301
·
Received October 21, 2014
Report
- Report Number
- 2183502-2014-00783
- Event Type
- Malfunction
- Date Received
- October 21, 2014
- Report Date
- October 20, 2014
- Manufacturer
- SMITHS MEDICAL MD, INC.
- Product Code
- FRN
- PMA / PMN Number
- K040899
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. A REVIEW OF THE EVENT HISTORY LOG CONFIRMED THE ERROR HAD OCCURRED. THE ERROR WAS DUPLICATED THROUGH TESTING. DURING VISUAL INSPECTION THE POSITION POTENTIOMETER AND SIZE POTENTIOMETER WERE FOUND WORN. BOTH COMPONENTS WERE REPLACED. AFTER REPAIR AND RECALIBRATION THE DEVICE WAS FOUND TO PASS ALL DELIVERY, ACCURACY AND FUNCTIONAL TESTS.
Description of Event or Problem · 1
A REPORT WAS RECEIVED THAT STATED THE PUMP ALARMED "CHECK CLUTCH." NO PATIENT INJURY OR ADVERSE EVENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 668255 | MEDFUSTION 3500 SYRINGE INFUSION PUMP | FRN - SYRINGE INFUSION PUMP | FRN | SMITHS MEDICAL MD, INC. | 3500 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |