FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4230060 · Received October 11, 2014

Report

Report Number
2032227-2014-37011
Event Type
Malfunction
Date Received
October 11, 2014
Date of Event
September 12, 2014
Report Date
September 12, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP POWERED UP PROPERLY AFTER BATTERY INSTALLATION. THE INSULIN PUMP PASSED THE SELF TEST, RESET ERROR TEST AND DISPLACEMENT TEST. NO UNEXPECTED RESET ERROR ALARMS WERE NOTED. THE INSULIN PUMP WAS RECEIVED WITH A CORRODED BATTERY TUBE. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND MINOR SCRATCHES ON THE DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT INSULIN PUMP ALARMED UNEXPECTED RESTART. CUSTOMER'S BLOOD GLUCOSE READING WAS 138 MG/DL. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
645697 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAP

Patients

Seq Age Sex Outcome Treatment
1 67 YR