FDA Adverse Event
Injury
Summary report: N
UNKNOWN
MDR report key: 4223930
·
Received October 30, 2014
Report
- Report Number
- 4223930
- Event Type
- Injury
- Date Received
- October 30, 2014
- Date of Event
- October 20, 2014
- Report Date
- October 30, 2014
- Manufacturer
- UNKNOWN
- Product Code
- FPO
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
RESIDENT ROLLED OFF SHOWER BED. LACERATION TO FOREHEAD, TO EMERGENCY ROOM FOR FEW SUTURES AND RETURNED TO FACILITY. UPON INVESTIGATION AFTER EVENT OCCURRED AND EQUIPMENT WAS INSPECTED, IT WAS DETERMINED THAT THE PIN WAS LOOSE AND NOT POSITIONED CORRECTLY IN LOCK OF SHOWER BED. PIN WAS NOT SECURELY INSERTED ALL THE WAY THROUGH. ALL SHOWER BEDS WERE INSPECTED BY MAINTENANCE AND PINS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 696530 | UNKNOWN | SHOWER BED TRANSPORT | FPO | UNKNOWN | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Other |