FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4223756
·
Received November 4, 2014
Report
- Report Number
- 2032227-2014-46624
- Event Type
- Malfunction
- Date Received
- November 4, 2014
- Date of Event
- September 11, 2014
- Report Date
- September 11, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UNIT RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED. UNIT RECEIVED WITH CRACKED CASE ON DISPLAY WINDOW CORNERS, MINOR SCRATCHED LCD WINDOW, CRACKED RESERVOIR TUBE LIP, AND CRACKED BELT CLIP SLOT. NO RESERVOIR COMPARTMENT ANOMALY WAS NOTED. NO MOISTURE DAMAGE ON MOTOR ASSEMBLY OR ELECTRONIC ASSEMBLY NOTED DURING VISUAL INSPECTION.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP AND STATED THAT CONDENSATION WAS VISIBLE IN THE DISPLAY. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 133 MG/DL. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 705478 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR |