FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4223557 · Received November 4, 2014

Report

Report Number
2032227-2014-47444
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
October 5, 2014
Report Date
October 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD A BLANK DISPLAY DUE TO A BROKEN SOLDER JOINT OF THE POWER CONNECTOR ON THE INTERFACE BOARD. THE INSULIN PUMP HAD A CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER'S MOTHER REPORTS THAT CUSTOMER HAS EXPERIENCED A BLANK DISPLAY SCREEN. CUSTOMER'S BLOOD GLUCOSE WAS 257 MG/DL. AFTER TROUBLESHOOTING, DISPLAY SCREEN DID NOT RETURN, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
707215 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LNAB

Patients

Seq Age Sex Outcome Treatment
1 8 YR