FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 4223535 · Received November 4, 2014

Report

Report Number
2955842-2014-05587
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
September 23, 2014
Report Date
October 10, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND BOTH PITCH AND UP AND DOWN CABLES BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. ISI HAS CONDUCTED A DEVICE HISTORY RECORD (DHR) REVIEW FOR THIS DEVICE AND DID NOT FIND ANY NON-CONFORMANCES THAT WERE RELATED TO THIS REPORTED EVENT. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN PITCH CABLES FOUND DURING FAILURE ANALYSIS COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SURGICAL PROCEDURE, A BROKEN CABLE WAS IDENTIFIED ON A PROGRASP FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED. THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO THE PATIENT.